Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.021 seconds
AbstractAbstract
[en] PIXE (Particle Induced X-ray Emission analysis) has been established as a method of trace element analysis. This work compares the detection limits using 2.13 and 2.55 MeV protons with those attainable using 4He2+ -ions of energies 6.07, 7.00, 8.52 and 9.12 MeV as projectile. This comparison was carried out using three common aerosol substrates: Kimfol, Nuclepore and Teflon. The detection limits using protons or 4He2+ -ions were of the same order but the lower detection limits with 4He2+ -ions according to theory were not observed, largely due to a higher intensity of gamma-quanta being Compton-scatterd in the Si(Li) detector. An improvement of the experimental chamber making it more suitable for 4He2+ -ions should lower the detection limits using 4He2+ as projectile. (author)
Original Title
En jaemfoerelse av detektionsgraenserna vid PIXE-spaaraemnesanalys mellan 2.13 och 2.55 MeV protoner och 6.07, 7.00, 8.52 och 9.12 MeV 4He2+ foer naagra typiska provsubstrat
Primary Subject
Source
Apr 1984; 53 p; Thesis (Civilingenjoer AV).
Record Type
Report
Literature Type
Thesis/Dissertation
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue