[en] A technical description is given of an apparatus for remote sensing of the activity distribution along a radioactive source, which is based on gamma scanning. Test results with extended sources are reported and recommendations are made on improving the equipment. The apparatus was tested with line sources of californium-252, curium-244, and americium-241. The amount of californium-252 was determined by a relative method involving comparison of the neutron count rates from the test source and a standard one. The amounts of curium and americium in the sources were determined by a radiometric method after dissolution
Source
Translated from Radiokhimiya; 29: No. 3, 409-411(May-Jun 1987).
Record Type
Journal Article
Literature Type
Translation
Journal
Country of publication
AMERICIUM 241, CADMIUM TELLURIDES, CALIFORNIUM 252, COLLIMATORS, COUNTING RATES, CURIUM 244, DISTRIBUTION, GAMMA DETECTION, ISOTOPE RATIO, NEUTRON DETECTION, PERFORMANCE TESTING, RADIOACTIVITY, RADIOISOTOPE SCANNERS, RADIOISOTOPE SCANNING, RELIABILITY, RESOLUTION, SEMICONDUCTOR DETECTORS, SENSITIVITY, SPONTANEOUS FISSION RADIOISOTO
ACTINIDE NUCLEI, ALPHA DECAY RADIOISOTOPES, AMERICIUM ISOTOPES, CADMIUM COMPOUNDS, CALIFORNIUM ISOTOPES, CHALCOGENIDES, COUNTING TECHNIQUES, CURIUM ISOTOPES, DETECTION, EVEN-EVEN NUCLEI, HEAVY NUCLEI, ISOTOPES, MEASURING INSTRUMENTS, NUCLEI, ODD-EVEN NUCLEI, RADIATION DETECTION, RADIATION DETECTORS, RADIOISOTOPES, TELLURIDES, TELLURIUM COMPOUNDS, TESTING, YEARS LIVING RADIOISOTOPES