Drozdowicz, K.; Hoek, M.; Aronsson, D.
Chalmers Univ. of Tech., Goeteborg (Sweden). Dept. of Reactor Physics1990
Chalmers Univ. of Tech., Goeteborg (Sweden). Dept. of Reactor Physics1990
AbstractAbstract
[en] The report describes measurements performed for the energy calibration of the TANSY neutron detectors (two arrays of 16 detectors each one). The calibration procedure determines four calibration parameters for each detector. Results of the calibration measurements are given and test measurements are presented. A relation of the neutron detector calibration parameters to producer's data for the photomulipliers is analysed. Also the tests necessary during normal operation of the TANSY neutron spectrometer are elaborated (passive and active tests). A method how to quickly get the calibration parameters for a spare detector in an array of the neutron detectors is included
Source
May 1990; 80 p
Record Type
Report
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume