[en] Moving beam technique in performing total body irradiation (TBI), which sweeps the X-ray beam completely over the whole body axis, is a useful technique for small treatment room in our hospital. However, homogenous dose distributions for patients cannot be obtained without changing the output during gantry rotation. We developed a method for obtaining a homogenous dose distribution by changing the software of the control unit of the linac without any change in the construction of the machine. The change of output/degree and/or the field width at each gantry angle can be automatically calculated by our method, the result, homogenous dose of ±3% at the reference plane was obtained in treating a useful length of 200 cm. From this study, though the results are still preliminary, our method might be promising in performing TBI. (author)
Primary Subject
Record Type
Journal Article
Journal
Country of publication
BEAMS, BODY, DISEASES, DISTRIBUTION, EXTERNAL IRRADIATION, HEMATOPOIETIC SYSTEM, IMMUNE SYSTEM DISEASES, IRRADIATION, LEPTON BEAMS, MEDICINE, MOCKUP, NEOPLASMS, ORGANS, PARTICLE BEAMS, RADIATION DOSE DISTRIBUTIONS, RESPIRATORY SYSTEM, SENSE ORGANS, SPATIAL DISTRIBUTION, STRUCTURAL MODELS, THERAPY, TISSUES