Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.023 seconds
Maeng, Seongjin; Lee, Sang Hoo; Kwon, Dahye; Seo, Jihye; Seo, Kyung Won
Proceedings of the KNS 2015 spring meeting2015
Proceedings of the KNS 2015 spring meeting2015
AbstractAbstract
[en] As the use of X-ray tubes is widely spread mainly for medical diagnostic purposes or industrial applications, there is increasing demand for accurate and convenient way getting of X-ray beam spectral information. While measurement methods may provide quite accurate spectral information, these methods still require expensive detectors (example: HPGe, High Purity Germanium detector) and some conversion of measurement information into real spectrum. It is concluded that Laplace transform-based spectral reconstruction technique given in equations (1) and (2) works well for a 50-kV X-ray source. In this paper we obtained the attenuation curve by the use of MCNPX simulations. We were able to rebuild the X-ray spectrum of 50 kV through this research by Monte Carlo simulation (fitting parameters, a: 1.2921, b: 0.2342, ν: 0.6190, R-squared: 0.9930)
Primary Subject
Secondary Subject
Source
Korean Nuclear Society, Daejeon (Korea, Republic of); [1 CD-ROM]; May 2015; [3 p.]; 2015 spring meeting of the KNS; Jeju (Korea, Republic of); 6-8 May 2015; Available from KNS, Daejeon (KR); 5 refs, 5 figs, 1 tab
Record Type
Miscellaneous
Literature Type
Conference
Country of publication
Reference NumberReference Number
Related RecordRelated Record
INIS VolumeINIS Volume
INIS IssueINIS Issue