Benlloch, J.M.; Carrilero, V.; Gonzalez, A.J.; Catret, J.; Lerche, Ch.W.; Abellan, D.; Garcia de Quiros, F.; Gimenez, M.; Modia, J.; Sanchez, F.; Pavon, N.; Ros, A.; Martinez, J.; Sebastia, A., E-mail: benlloch@ific.uv.es2007
AbstractAbstract
[en] We have constructed a small animal PET with four identical detector modules, each consisting of a continuous LYSO crystal attached to a Position Sensitive Photomultiplier Tube (PSPMT). The dimensions of the continuous crystal are 50x50 mm2 and 10 mm thickness. The modules are separated 11 cm between each other in the scanner. In this paper we discuss the method used for the calibration of the camera for this special system with continuous detectors. We also present the preliminary values for the main performance parameters such as spatial and energy resolution, and sensitivity of the system
Primary Subject
Source
EuroMedIm 2006: 1. international conference on molecular imaging technology; Marseille (France); 9-12 May 2006; S0168-9002(06)01819-5; Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Literature Type
Conference
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002; ; CODEN NIMAER; v. 571(1-2); p. 26-29