AbstractAbstract
[en] Electrochemiluminescene immunoassay (ECLIA) and enzyme-linked immunosorbent (ELISA) were used to detect the different concentrations of serum HBsAb, in order to compare the results of ECLIA and ELISA. The result showed that intra-assay coefficient of variation of ECLIA was about 0.95% for high value, 1.13% for mean values and 2.63% for low value, while that of ELISA was about 5.76%, 12.8% and 15.9%, respectively. The interassay coefficient of ECLIA was about 4.03% for high values, 4.93% for mean values and 7.34% for low values, while that of ELISA was about 10.1%, 19.6% and 25.2%, respectively. The analytical sensitivity of ECLIA was about 4.0IU/L, while that of ELISA is about 19.0IU/L. Only in 3 samples, the results measured by ECLIA and ELISA were different (ECLIA: positive; ELISA: negative) among 165 samples. It is concluded that the met hod of ECLIA is more efficient than ELISA for detection of HBsAb in serum. (authors)
Primary Subject
Source
2 tabs., 2 refs.
Record Type
Journal Article
Journal
Labeled Immunoassays and Clinical Medicine; ISSN 1006-1703; ; v. 12(1); p. 41-42
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue