AbstractAbstract
[en] The aim of this study is to investigate the influence on image contrast, tube load and effective dose in paediatric fluoroscopy using added filtration. A Philips Omnidiagnosti is used in Nino Jesus Hospital and was therefore chosen for radiation dose measurements. The phantom consisted of varying methacrylate thickness to represent different patients sizes. All measurements were performed in automatic mode. For each exposure and additional filtration added the following data was recorded: tube voltage, tube current, air kerma rate on phantom surface and brightness on the image monitor. An 2026 electrometer (Radical Corporation) in conjunction with a model 2025-60 ion chamber were used for entrance dose measurements and the luxometer IL-400A (International Light) was used for brightness measurements on the image monitor: Evaluation of image quality was performed using a Leeds TOR TVF test object. Finally, the program PCXMC 1.5 based on the Monte Carlo method was used for calculating organ doses and the effective dose in fluoroscopy examinations. By increasing the filtration of the x-ray tube 1st Entrance radiation exposure can be decreased 58%, organ dose up to 40%, and effective dose up to 44%. 2nd The tube load increased up to 33%. 3rd Significant dose reduction is achievable without compromising image quality. The use of additional filtration in paediatric fluoroscopy should be evaluated taking into account dose reduction, additional tube loading and the possibility of some deterioration in image quality. (Author)
Original Title
Reduccion de dosis en exploraciones complejas mediante el uso de filtracion adicional en pacientes pediatricos
Primary Subject
Record Type
Journal Article
Journal
Radioproteccion (Madrid); CODEN RDPREY; v. 49(XIII); p. 95-98
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue
AbstractAbstract
[en] A female employee working in diagnostic radiology should take additional controls to protect the unborn child from ionizing radiations. The fetus is particularly sensitive to the effects of x-rays and, so, the determination of the equivalent dose to the unborn child is of interest for risk estimates from occupational exposures of the pregnant workers. The ian of this study is to develop a method for fetus dose estimate of a pregnant worker who participates in interventional radiology procedures. Factors for converting dosemeter readings to equivalent dose to the fetus have been measured using thermoluminescence dosimetry. Equivalent dose to the uterus is used to simulate the equivalent dose to the fetus during the first two months of pregnancy. Measurements at different depths are made to consider the variations in the position of the uterus between pregnant women. The normalized doses obtained are dependent on the beam quality. Accurate estimation of fetus doses due to occupational exposures can be made using the data provided in the current study. (Author)
Original Title
Estimacion de la dosis equivalente en feto durante la exposicion ocupacional de trabajadores gestantes en radiologia intervencionista
Primary Subject
Record Type
Journal Article
Journal
Radioproteccion (Madrid); CODEN RDPREY; v. 49(XIII); p. 200-203
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue
AbstractAbstract
[en] The aim of this study is to evaluate the efficiency of bismuth breast shield in CT chest paediatric examinations when automatic exposure control techniques are used. The influence in the radiation dose and the image quality has been evaluated with and without the breast shield in the scoutview. In addition the radiation dose in shielded and non-shielded areas has been compared. Measurements were made in a 16 cm diameter cylindrical PMMA phantom simulating a newborn, and older children were simulated controlling the maximum intensities allowed by the automatic exposure control system AutomA. The highest dose reduction (59%) was obtained with AutomA system and when the breast shield is not used in the scoutview. This reduction in the radiation dose does not mean a significant increase of noise level. The use of the breast shield in the scoutview yielded an increase in the radiation dose in non-shielded areas. The use of bismuth breast shield is recommended only after the scoutview in order to optimise the radiation dose in CT chest paediatric examination when using automatic exposure control AutomA. (Author)
Original Title
Valoracion en el paciente pediatrico de la eficacia de los protectores de bien procedimientos de TC de torax con sistemas de control automatico de intensidad
Primary Subject
Record Type
Journal Article
Journal
Radioproteccion (Madrid); CODEN RDPREY; v. 49(XIII); p. 80-83
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue