[en] The aim of this study is to evaluate and analyze the on position leakage in-patient plane and radiation safety compliance of telegamma unit over the period of 36-months. RMM (Roentgen per minute at 1 meter) measurements and leakage radiation measurements in-patient plane are the part of the Quality assurance programme. Both measurements were repeated at regular intervals. These recorded values have been taken up to assess the shielding adequacy of source housing of telecobalt unit over the period of years. Measured values at 69.0 cm and 156.0 cm from the beam axis 8 points located in-patient plane ranges from 0.027% to 0.081% and 0.0045% to 0.0062% respectively, which were well below the specified tolerance limit of 0.1% RMM
Primary Subject
Source
AOCMP-AMPICON 2017: 17. Asia-Oceania Congress of Medical Physics; Jaipur (India); 4-7 Nov 2017; 38. annual conference of Association of Medical Physicists of India; Jaipur (India); 4-7 Nov 2017; 1 fig., 1 tab.
Record Type
Journal Article
Literature Type
Conference
Journal
Journal of Medical Physics; CODEN JMPHFE; v. 42(suppl.1); p. 225-226
Country of publication
BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, COBALT ISOTOPES, DISEASES, DOSES, INTERMEDIATE MASS NUCLEI, INTERNAL CONVERSION RADIOISOTOPES, ISOMERIC TRANSITION ISOTOPES, ISOTOPES, MANAGEMENT, MEDICINE, MINUTES LIVING RADIOISOTOPES, NUCLEAR MEDICINE, NUCLEI, ODD-ODD NUCLEI, RADIOISOTOPES, RADIOLOGY, THERAPY, YEARS LIVING RADIOISOTOPES
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue