FOMA: A Fast Optical Multichannel Analyzer
Haskovec, J.S.; Bramson, G.; Brooks, N.H.; Perry, M.
General Atomics, San Diego, CA (USA)1989
General Atomics, San Diego, CA (USA)1989
AbstractAbstract
[en] A Fast Optical Multichannel Analyzer (FOMA) was built for spectroscopic measurements with fast time resolution on the DIII-D tokamak. The FOMA utilizes a linear photodiode array (RETICON RL 1024 SA) as the detector sensor. An external recharge switch and ultrafast operational amplifiers permit a readout time per pixel of 300 ns. In conjunction with standard CAMAC digitizer and timing modules, a readout time of 500 μs is achieved for the full 1024-element array. Data acquired in bench tests and in actual spectroscopic measurements on the DIII-D tokamak is presented to illustrate the camera's capability. 4 refs., 3 figs
Primary Subject
Source
Dec 1989; 3 p; 13. IEEE symposium on fusion engineering; Knoxville, TN (USA); 2-6 Oct 1989; CONF-891007--112; CONTRACT AC03-89ER51114; NTIS, PC A02/MF A01 as DE90004093; OSTI; INIS; US Govt. Printing Office Dep