Preparation of 82Sr/82Rb Generator and Positron Emission Tomographic Image of Normal Volunteer
AbstractAbstract
[en] A 82Sr/82Rb generator was prepared by loading 82Sr to preconditioned tin dioxide column. The generator was eluted by normal saline with flow rate up to 8 ml/min, and the eluted radioactivity was monitored by dose calibrator. Radioactivity began to come out at 5 ml and reached to peak around 9 ml. The total eluted radioactivity increased linearly with flow rate, and the maximum obtained radioactivity was 35mCi at 8 ml/min. The 82Rb preparation was proven to be free from both strontium radioactivity and pyrogen. The 82Rb was injected to normal female volunteer and positron emission tomographic image of heart was obtained successfully.
Primary Subject
Source
15 refs, 5 figs
Record Type
Journal Article
Journal
Journal of the Korean Radiological Society; ISSN 1738-2637; ; v. 28(3); p. 326-330
Country of publication
Descriptors (DEI)
Descriptors (DEC)
ALKALINE EARTH ISOTOPES, BETA DECAY RADIOISOTOPES, BETA-PLUS DECAY RADIOISOTOPES, COMPUTERIZED TOMOGRAPHY, DAYS LIVING RADIOISOTOPES, DIAGNOSTIC TECHNIQUES, ELECTRON CAPTURE RADIOISOTOPES, EMISSION COMPUTED TOMOGRAPHY, EVEN-EVEN NUCLEI, HOURS LIVING RADIOISOTOPES, INTERMEDIATE MASS NUCLEI, ISOTOPES, MINUTES LIVING RADIOISOTOPES, NUCLEI, ODD-ODD NUCLEI, RADIOISOTOPES, RUBIDIUM ISOTOPES, STRONTIUM ISOTOPES, TOMOGRAPHY
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue