Measures to reduce minimize variation in steel analysis by Optical emission spectrometer (OES)

Measures to reduce minimize variation in steel analysis by Optical emission spectrometer (OES)

The variation in analysis of steel using an Optical Emission Spectrometer (OES) can occur due to several factors related to the sample, equipment, or environment. Below are common reasons for variation, along with precautions to reduce the inconsistencies:

Reasons for Variation in OES Analysis:

  1. Surface Contamination: Presence of oils, oxides, rust, or dirt on the sample surface can lead to inaccurate readings.
  2. Improper Sample Preparation: Uneven surface finish, insufficient grinding, or polishing can affect the emission signal and lead to inconsistent results.
  3. Electrode Condition: Worn or contaminated electrodes can cause variation in the spark, impacting the quality of the emission spectrum.
  4. Instrument Calibration: Calibration errors or drift over time can cause shifts in the analytical results.
  5. Matrix Effects: Variations in the chemical composition of the steel matrix (presence of trace elements or impurities) can influence the emission of certain elements, leading to errors.
  6. Temperature Fluctuations: Changes in room or sample temperature can affect the sensitivity of the equipment and the analysis.
  7. Inconsistent Argon Purging: Argon is used to protect the spark environment from contamination by atmospheric gases. Any inconsistency in argon flow or purity can lead to inaccurate results.
  8. Instrumental Factors: Aging components, such as detectors, optical fibers, or light sources, may degrade performance over time.
  9. Operator Error: Improper handling of samples or incorrect instrument settings can result in analytical variation.

Precautions to Reduce Variation in OES Analysis:

  1. Proper Sample Preparation: Ensure the sample surface is clean, smooth, and flat. Grinding and polishing should be consistent, with the removal of any surface oxides or contaminants. Use a standardized procedure for all samples to maintain consistency. Use same grit for polishing CEM / standard samples and samples to be analysed.
  2. Clean Electrode and Sample Stand: Regularly clean and maintain the electrodes and the sample stand to avoid contamination and ensure a consistent spark.
  3. Regular Calibration and Drift Correction: Calibrate the OES with certified reference materials (CRMs) regularly. Perform drift correction during long analysis runs to adjust for any instrument drift over time.
  4. Consistent Argon Supply: Ensure high purity and steady flow of argon gas during the analysis. Regularly check for leaks in the argon supply lines.
  5. Control Room Temperature: Maintain a stable temperature in the room where the OES is installed to minimize the impact of thermal changes on the instrument.
  6. Regular Instrument Maintenance: Follow the manufacturer's maintenance schedule, including the replacement of worn components (e.g., electrodes, spark stand, detectors) to keep the instrument in optimal condition.
  7. Train Operators: Ensure operators are well-trained on the correct procedures for sample preparation, instrument handling, and analysis methods to minimize human error.
  8. Use Standardized Sample Location: Always spark in the same area of the sample and ensure multiple sparks are performed for more consistent results.

By following these precautions, the variation in OES analysis can be minimized, leading to more accurate and reliable results for steel analysis.