Understanding Transfer Path Analysis - Part 4: Practical Advice
Sensor Technology Microphones
Microphone Mounting
Microphones are used to record required airborne sound signals. Microphone stands, magnets or screw connections are suitable for fastening the microphones. In order to avoid structure-borne noise-induced crosstalk, it is not recommended to attach the microphones directly to the source or test object. If this cannot be avoided, decoupling elements such as foams must be used.
Positioning of the Microphone
In order to determine the sound pressure in the near field of the source, the microphones are to be positioned at a distance of about 10–15 cm from the surface of the source. Concerning measurements at flow openings (e.g., intake, and exhaust tailpipe), the microphone must not be positioned directly in the air flow. In practice, alignment at a 45° angle to the opening has proven effective.
Microphone Diameter
The most commonly used microphones in TPA are sound pressure microphones with omni-directional characteristics. The diameter of the microphones used is usually ¼ or ½ inch. ¼‘‘ microphones are capable of recording signals with a good signal-to-noise ratio. ½‘‘ microphones are used if a higher signal-to-noise ratio is required for subsequent evaluations.
However, ½‘‘ microphones are not capable of recording as high sound pressures as can ¼‘‘ microphones.
As a consequence, the selection of the microphone depends on the task.
Accelerometers
IEPE (Integrated Electronics Piezo Electric)
Structure-borne sound signals are often recorded with accelerometers. In most cases, piezoelectric accelerometers are used for this purpose. With piezoelectric accelerometers, an electric charge is generated by the acceleration of a mass coupled to a piezo element. Using IEPE technology(Integrated Electronics Piezo Electric) reduces electrical interference susceptibility, so high signal quality can be achieved even with long cables to the sensor.
Charge "Type" Sensors
Depending on the field of application, charge-capacitive sensors (charge sensors) are also used as an alternative. Charge sensors separate the role of the sensor from that of the signal amplifier. The advantage of these sensors is that the amplifiers can be configured in a more complex way. The disadvantage is, however, that long cables may be required to transport the sensor signal to the amplifier.
In cases of extreme environments, the Charge "Type" Sensor may be required, but the consequence to this is additional cost in hardware, configuration, and potential error points.
Mounting
Accelerometers can be attached in various ways, e.g., with magnets, wax, adhesive, or screw connections. The choice of coupling directly affects the frequency range that can be measured with the sensor. For a durable, temperature-resistant connection in TPA measurements, the accelerometers are usually glued to the measurement object.
When attaching the sensors, it must be ensured that they are electrically decoupled from the measurement object.
In order not to damage the glued sensors after the measurement, they need to be rotated for removal. All sensors that are suitable for adhesive mounting have a hexagonal base plate so that an open-end wrench can be used to detach them if necessary.
Determining Structure-borne Sound Transfer Functions
Impact Hammer and Shaker
In most cases, the static structure-borne sound transfer functions are determined using an impact hammer or shaker. When using an impact hammer, force pulses are applied into the structure via hammer stroke in order to excite it over a broad band. An impact hammer has a force sensor that measures the force applied during the stroke. Shakers usually consist of a vibration source and a force sensor to record the applied force. For excitation, the shaker system is fed with an excitation signal. Both types of excitation feature specific advantages and disadvantages:
Measuring transfer functions using an impact hammer
Measuring with an Impact Hammer
The range of forces that can be applied to a structure depends on the specific structure at the point of impact and the strength of force application. The following applies: The shorter the impulse applied, the wider the frequency spectrum of the force.
Using Window Functions
In order to avoid nonlinearities, force application should be as small as possible. If this makes the signal-to-noise ratio too small, force application needs to be increased or a more sensitive sensor must be used. If the signal of the force excitation is noisy, this will reduce the significance of the transfer function. The noise component can be minimized by using a window function that either allows the force pulse to pass or otherwise outputs the value 0. In order to reduce the leakage effect, the window function must decay to 0 at the end.
Spurious Noise in the Response Signal
Even the system response to the force pulse, i.e., the acceleration signals, may be contaminated with noise. It is especially during the decay of the structure after impulse excitation that the superposition of the system response and noise make it difficult to distinguish the former. Spurious noise in the response signals often occurs when the force application point and the measurement point are far apart. Using suitable sensor technology and appropriate window functions will help to significantly reduce the noise component in the response signals. Typically, windows based on exponential functions are used.
Parameters for the Measurement
Important parameters for the impact hammer measurement:
To ensure that the beginning of the force impulse is not cut off, it is essential to use a pre-trigger (5-10% of the recording duration).
For this reason, it is advisable to start by recording a few untriggered test strokes and analyzing the decay times of the system responses.
Averaging reduces uncorrelated noise and provides statistical reliability to the results.
Improving Coherence
A meaningful transfer function can only be determined if sufficient energy was applied by the impact hammer and the measurement shows a high coherence between excitation and system response in the relevant frequency domain. The following measures help to reduce dips in coherence:
If there are only isolated dips in coherence in the force spectrum, they are usually caused by anti-resonances of the structure.
Differences of the Hammer Tips
Recommended by LinkedIn
If the force spectrum shows no excitation in the relevant frequency range, this may possibly be changed by selecting a different hammer tip. The following applies:
Double Stroke
If the force spectrum is not sufficiently constant in the relevant frequency domain, it may be assumed that more than one impulse was recorded per analysis block. If this is the case, the measurement needs to be repeated.
Spectrum with High Frequencies
As a rule, the force spectrum is meant to fall off at high frequencies. If this is not the case, the system was probably excited with excessive high frequencies, which may result in nonlinearities. However, the force spectrum should not fall off too much, as otherwise the noise floor will mask the force signal.
Measuring Transfer Functions using a Shaker
Correct Use of a Shaker
There are also some points to consider when measuring transfer functions with shaker excitation. For example:
If a large shaker system is not suitable to be used due to lack of space or with very lightweight structures, the Qlws Lightweight Shaker from Qsources can be used to determine the transfer functions.
Signals for Shaker Excitation
For excitation, the shaker system is fed with an excitation signal. The following signals are suitable for determining the transfer function using a shaker:
Using Window Functions
Depending on the signal type, the recorded time signal of the shaker must be windowed in order to improve the signal quality.
Transfer Function Matrix
Position of the Force Application
As described above, defined forces are applied into the structure using an impact hammer or shaker to determine the structure-borne sound transfer functions. The position of this force application is supposed to be as close as possible to the actual position of the force application in the operating case. The better the points match, the more accurately the force can be described during operation.
Overdetermined Matrices
For the matrix inversion method, a matrix with good matrix conditioning is crucial. In order for the matrix conditioning to be optimized, it can be overdetermined and/or regularized. Overdetermination requires accelerations to be determined at more response positions than are needed to describe the system. A common guideline is to have twice as many response than excitation points.
Performing Operational Measurements
Operational Measurements with an Active Source
Operational measurements are performed using an active source (e.g., on a test site or test bench). All airborne and structure-borne sound channels are recorded synchronously. This allows the phase relationship between the different paths to be determined and taken into account. In addition, other measured quantities, such as speed, torque, and CAN, can be recorded.
Selection of the Operating Conditions
When performing the TPA in the time domain, there is no restriction to stationary operating conditions. The source is to be operated in any critical operating condition and the resulting sound pressures and accelerations are to be recorded. Critical operating conditions are usually those in which disturbing noise occurs.
Possible Sources of Error
Naming of the Recordings
The amount of data that must be measured, analyzed and organized during TPA is very large. In order not to make any mistakes during evaluation, it must be ensured that the individual measurements can be reliably distinguished and assigned to the correct measurement points. The TPA Project in ArtemiS SUITE assists users with this task.
Systematic Errors
Systematic errors in TPA may be caused by the following:
Engineering Services of HEAD acoustics
This application note can of course only list a selection of practical tips and possible sources of errors. Thus, it only provides a brief insight into the subject. The Engineering Services of HEAD acoustics can provide you with more comprehensive support for your TPA projects. Contact our expert to benefit from our many years of experience and expertise: engineering@HEAD-acoustics.com
Stay Tuned for Part 5!
You can find the PDF version of this here on the HEAD acoustics website!
Do you have any questions? Your feedback is appreciated!
For questions on the content of this Article: BCremeans@HEADacoustics.com
For questions of the linked Document: Imke.Hauswirth@head-acoustics.com
For technical questions on our products: SVP-Support@head-acoustics.com
Sales & Application Engineer at HEAD acoustics, Inc.
5moSo many different ways in which you can use a TPA project. Test your mounts for damping properties with Effective Mount Transfer Matrix (EMTF), hone in on what occupants hear by doing BTPA (Binaural transfer path analysis). Use a combination of these and more to get an accurate fully interactive model!
Analysis, Know-how, Teamwork, and Solution - how to guarantee success! This article discuss all of those, and even what to watch out for.
Consultant - Engineering Services at HEAD Acoustics, Inc.
5moKnowing your test setup and what bad data looks like is the first step to being able to gather usable information. This is a great overview to get started! 🔨
I solve Audio/Acoustic challenges
6moSo many good details in this article! 👍 This is TPA 101.