Parametrisation of the collimator scatter correction factors of square and rectangular photon beams
AbstractAbstract
[en] Collimator scatter correction factors Sc have been measured with a cylindrical mini-phantom for five types of dual photon energy accelerators with energies between 6 and 25 MV. Using these Sc-data three methods to parametrize Sc of square fields have been compared including a third-order polynomial of the natural logarithm of the fieldsize normalised by the fieldsize of 10 cm2. Also six methods to calculate Sc of rectangular fields have been compared including a new one which determines the equivalent fieldsize by extending Sterling's method. The deviation between measured and calculated Sc for every accelerator, energy and all methods are determined resulting in the maximum and average deviation per method. Applied to square fields the maximum and average deviation were for the method of Chen 0.64% and 0.15%, of Szymzcyk 0.98% and 0.21%, and of this work 0.41% and 0.10%. For the rectangular fields the deviations were for the method of Sterling 1.89% and 0.50%, of Vadash 1.60% and 0.28%, of Szymczyk et al. 1.21% and 0.25%, of Chen 1.84% and 0.31% and of this work 0.79% and 0.20%. Finally, a recommendation is given how to limit the number of fields at which Sc should be measured